| Model name | SX-700E | ||
|---|---|---|---|
| Operating temperature range | Sterilizing | 105 ∼ 135°C (0.019 ∼ 0.212MPa) | |
| Heating | 45 ∼ 104°C (0 ∼ 0.015MPa) | ||
| Warming | 45 ∼ 95°C | ||
| Maximum operating pressure | 0.25MPa | ||
| Temperature | Display | Digital | |
| Pressure gauge | Display | Analog | |
| Display range | 0 - 0.4MPa | ||
| Heat source | 3.0kW electric heater | ||
| Safety device | • Water level sensor, • Leakage breaker, • Lid interlock, • Over-heat prevention, • Over-pressure prevention, • Temperature sensor break detection, • Safety valve | ||
| Protection type against electrical shock | Class I equipment | ||
| Time | Display | Digital | |
| Display range |
Sterilizing | 0:00 - 9:59, 0 - 99hours, 0 - 999min.(0:00 - 9:59,10 - 99hours changeable) | |
| Heating | |||
| Warming | Variable: 0 - 99hours | ||
| Pressure chamber type | PED Category II | ||
| Chamber dimensions | φ370 x 774mm | ||
| Chamber capacity | Effective internal volume:69L, Internal volume:79L |
||
| Chamber material | SUS304 | ||
| Dimensions (mm) | 470W x 528D x 1003Hmm (With projection:625D) |
||
| Weight | 72kg | ||
| Rated voltage | 230V AC | ||
| Power input | 13A | ||
| Required power supply | Single-phase 230V AC (50/60 Hz)15 A or more |
||
| Power consumption (calorific value) | 3.0kW(2580kcal/h) | ||
| Environmental Conditions | When opereting the autoclave, observe the environmental conditions given below. Ambient temperature: 10 to 35°C Almospheric pressure: 860 to 1060hPa Relative humidity: 30 to 85% Maximum gradient: 2 |
||
| Accessories | Stainless baskets 2 (φ345×181 mm) |
||
| • Caster stopper 4, • Chamber bottom Plate 1, • Operator's Manual 1, • Clear Plastic Accessory Case 1, • Attachment Screw for Clear Plastic Accessory Case 1, • Inspection Sheet 1, • Warranty card 1, • Customer card 1 | |||