Power
Power supply: AC input without external power
adaptor or battery operation
AC power: 100 to 240 VAC ± 10% 60 VA
Battery-type: rechargeable Lithium ion battery
11.1 V typical, 2,500 mAh
- capacity: 6 hours of continuous operation
without recording or 500 ECGs in
2.5 x 4 format at 25 mm/s and 10
mm/mV
- charge time: 3.5 hours with power off
Processing
ECG analysis
samplig rate: 500 samples/second (sps)
Digital
sampling rate: 1,000 samples/second/channel
Pacer detection
sampling rate: 16,000 samples/second/channel
ECG amplifier: DC-coupled
Acquisition mode: Pre-acquisition or post-acquisition
provide 10 seconds of
instantaneous ECG acquisition
Baseline drift filter: 0.05 Hz, Baseline Drift Removal
(BDR)
AC filter: 50/60 Hz ± 0.3 Hz
ADC: 24 bit
Imput impedance: 50 MD at 10 Hz, defibrillator
protected
Patient leakage: < 10 μA
Heart rate meter: 30 to 300 BPM ± 10% or ± 5
BPM, whichever is greater
Sensitivity/gain: 5, 10, 20 mm/mV, Auto
DIsplay 24 bit Colour display
TFT LCD with LED graphics backlit
Resolution: 800x400
Display data: Patient ID, gender, age, heart rate,
clock, battery power indicator,
waveforms, lead labels, speed,
gain, filter settings, warning
messages, information messages,
network status and USB
Thermal printer: Dot array/Width 80 mm/speed 5,
12.5, 25, 50 mm/s
Software
Measurement and interpretation:
according to University of
Glasgow ECG analysis algorithm
Resting ECG mode: Records and prints 12-lead
resting ECG with 10-second
duration as a standard
Patient information: Patient ID, secondary ID, age,
date of birth, gender, race, medication
class and V3 electrode
placement
Internal storage: 800 ECGs in internal memory
ECG Storage format:
PDF format and Mindray storage
format
Multi-language internal software:
GB, FR, IT, ES, DE, PL, PT, TR, RU,
HU, RO, CZ, CN
Report formats Thermal printer/PDF/A4/letter
Size/Weight: 250x194xh 56 mm/1.2 kg
including battery